MIDS-20X mutual inductance
detection system
- dedicated, ultra-low noise mutual inductance
detection system for measuring the SRD1000 and CMN1000
sensors;
- design ensures minimal rf-heating
of ultra-low temperature experiments;
- (remotely) adjustable
preamplifiers are positioned close to cryostat, the control
unit is near the user for optimal signal performance and
ease of use;
- 'plug and play': sensors and
electronics are matched, no adjustments are required to
measure the full range 10 mK to 10 K;
- primary currents sensors: 50
or 125 µA @
976.5 Hz;
- DC system output voltage: 0 -
2 V, proportional to the sensor signal level;
- two 24-bit ADC's with optical
USB interface (galvanically isolated from the system) to
sample the output signals;
- system comes with
software tools (DLL interface and LabVIEW VI) for data
acquisition and processing;
- overall temperature
coefficient of the electronics is better than 50 ppm/°C;
- the MIDS-202 is a 2-channel
detection system for simultaneous measurement of an SRD1000
and CMN1000 sensor;
- the MIDS-201 is a 1-channel
version of the MIDS-202, an optional multiplexer unit
(SSB-01) allows
switching between 2 measurement channels.
|